Product description:
The VEGAPULS 21/31 are radar level sensors for continuous level measuring.
Main characteristics:
Operating voltage: 12 ... 35 V DC (24 V DC nom.)
Output signal: two-wire 4…20mA /HART
Ambient temperature range: -40°C to 70°C
Degree of protection: IP66/67
Certificates for use in hazardous areas*:
IECEx KIWA 19.0015X:
Type of protection: Ex ia
Process characteristics:
Max. measuring range: up to 15 m
Max. accuracy: ±2 mm
Max. process pressure: -1...+3 bar
Process temp.: -40°C ... +80°C
Additional place of manufacturing:
VEGA Americas
3877 Mason Research Pkwy 45036
Mason, OH
United States of America
Documents:
VEGAZW-6-64984
Drawings: BB1512-1; BB1540-3; GE4049,01; LP1512-1; LP1540-3; SB1512-1; SB1540-3;
Test reports:
Hansecontrol Zertifizierungsgesellschaft mbH: - 2001-0061601R01B_EMC;
Paconsult: - 20-12891A-BE-VEGAPULS (VEGA);
Remarks:
- The products fulfill EC-Code: 31/41.
- The EMC requirements for installation on the Bridge and Deck Zones will be fulfilled by the equipment.
- For Passenger Vessels, level switches are not to be installed on flammable fluid tanks where they require tank wall penetration below the tank top.
- The sensing device is to be fitted with a valve or arranged such that it may be isolated and removed without emptying the tank or without impairing the integrity of pressurized system.
- Each particular application and installation is to be specifically approved in conjunction with the relevant system and tank arrangements.
- Wireless data communication used for vessel services (such as essential services, category I/II/III systems, etc.) will be specially considered provided the use of the wireless data communication results in a improvement in the safety of the vessel, compared to wired data communications.
* For installation in hazardous areas the requirements of the applicable safety certificates are to be complied with.