Product Description:
This system is a continuous gas monitoring system. The modular design of the system makes it possible to monitor different gases according to customer specification. It can monitor the SO2 and CO2 concentrations in exhaust gases. It provides an accurate measurement of SO2 in ppm, CO2 in percent as well as presents the SO2/CO2 ratio. The gas analyzer is based on both non-dispersive infrared (NDIR) principle and non-dispersive Ultraviolet (NDUV) principle. These principles makes it possible to detect a wide range of gasses among others SO2, CO2, NO2, NO, CH4, CO, HC, N2O and H20. The system includes heated probe(s) and heated hose(s) to maintain the extractive exhaust gas sample at a sufficient temperature to avoid condensed water and loss of SO2.
Technical Data:
G7200 Multigas Monitoring System |
Type | G7200 |
Function | Monitoring of CO2, SO2, NO2, NO, CH4, CO, N2O, HC, H2O
and other relevant gasses. |
Equipment including | Analyzer modules, gas cooler, gas pump (s), condensate
pump(s), gas filtration system, flow transmitter, solenoid
valves, air filter regulator and oil/particle separator, air dryer,
HMI, PLC. |
Number of Gas Analyzer Modules | 1-3 depending on configuration |
Power Supply | 230 VAC – 50/60 Hz |
Power Consumption | 16 A - 40 A dependent on system configuration – refer to the
electrical drawings of each system |
External Communication | Modbus TCP/IP (RJ45).
Optional Modbus RTU |
Air supply quality | ISO 8573-1 Class 3
Free from traces of measured gasses. |
Sample flow | 0.33 – 1.67 l/min.
Default 0,5L/min. |
Max ambient temperature | Class A
From 5°C to 45 °C |
Humidity | RH up to 96% at all relevant temperatures |
Vibration | Class A |
EMC | Class A |
Material/Enclosure Class | Painted mild steel RAL 7035
IP54 |
Dimensions/Weight | 1200 x 800 x 300 mm (H x W x D)
Approx. 118 kg depending on configuration |
Optional Modules |
IoT module | Establish a remote connection link between system and user |
Communication Converter Module | Converts bus communication to other standard bus
communication protocols |
Ethernet Switch | Used to connect optional modules |
TYPE APPROVAL CERTIFICATE
No. MAC045724XP/001
Enclosure Page 2 of 3
G7200
Analyzer Modules:
Type G7010 |
Gas Analyzer Measuring principle | NDIR |
Measuring Gasses and typical Ranges | CO2: 0 – 10 %
SO2: 0 – 200 ppm
Other ranges upon request |
Accuracy/Linearity | ≤ ±2 % of reading or
≤ 0.3 % of full scale whichever is larger (except zero) |
Precision/Repeatability | ≤ ±1 % of full scale above 100 ppm or
≤ ±2 % of full scale below 100 ppm |
Noise | ≤ ±2 % of full scale |
Zero drift | < ±2 % of full scale |
Span drift | < ±2 % of full scale |
Calibration | Zero Calibration: Automatic using compressed air
Span Calibration: Automatic using inbuilt cuvette which hold
span gas and also possibility to connect mixed test gases for
verification |
Type G7210 |
Gas Analyzer Measuring Principle | NDIR |
Measuring Gasses and typical Ranges | Gas Component | Full range | Default range |
NO CH4 H2O
CO2 CO N2O
HC | 0 – 5000 ppm
0 – 100 %
0 – 5%
0 – 100 %
0 – 100 %
0 – 100 %
0 – 100 % | 0 – 1000 ppm
0 – 5000 ppm
0 – 20000 ppm
0 – 20 %
0 – 5000 ppm
0 – 2000 ppm
0 – 5000 ppm |
Note: Combination of gasses and ranges are hardware
dependent. Each module can be configured for maximum 4
individual gasses. Kindly contact us for specific requirements. |
Accuracy/Linearity | ≤ ±2 % of reading or
<= 0.3 % of full scale whichever is larger (except zero) |
Precision/Repeatability | ≤ ±1 % of full scale above 100 ppm or
≤ ±2 % of full scale below 100 ppm |
Noise | ≤ ±2 % of full scale |
Zero drift | < ±2 % of full scale |
Span drift | < ±2 % of full scale |
Calibration | Zero Calibration: Automatic using compressed air
Span Calibration: Manually by connecting mixed test gases for
verification |
Compensations | Fully temperature and pressure compensated |
TYPE APPROVAL CERTIFICATE
No. MAC045724XP/001
Enclosure Page 3 of 3
G7200
Type G7230 |
Gas Analyzer Measuring Principle | NDUV + NDIR |
Measuring Gasses and typical
Ranges | Gas
Component | Full range | Default range |
NO
CH4
H2O
CO2
CO
N2O
HC
SO2
NO2 | 0 – 5000 ppm
0 – 100 %
0 – 5%
0 – 100 %
0 – 100 %
0 – 100 %
0 – 100 %
0 – 10 %
0 – 5000 ppm | 0 – 1000 ppm
0 – 5000 ppm
0 – 20000 ppm
0 – 20 %
0 – 5000 ppm
0 – 2000 ppm
0 – 5000 ppm
0 – 300 ppm
0 – 100 ppm |
Accuracy/Linearity | ≤ ±2 % of reading or
<= 0.3 % of full scale whichever is larger (except zero) |
Precision/Repeatability | ≤ ±1 % of full scale above 100 ppm or
≤ ±2 % of full scale below 100 ppm |
Noise | ≤ ±2 % of full scale |
Zero drift | < ±2 % of full scale |
Span drift | < ±2 % of full scale |
Calibration | Zero Calibration: Automatic using compressed air
Span Calibration: Manually by connecting mixed test gases for
verification |
Compensations | Fully temperature and pressure compensated |
Reference Documents:
P&I & Arrangement & Installation Drawings + System Description filed under RINA dwg no. PLMC-10889
Components + Electrical Drawings + IP Rating Assessment + SQAP filed under RINA dwg no. PLMC-10888
Test Reports + Control Plan + User Manual + Certificates filed under RINA dwg no. PLMC-10890
PLMC13414/ PLMC 13415 + G7200 Hazid evaluation for dual fuel LNG engines
Application Fields:
The G7200 Multigas Monitoring System may be used on board ships and other units classed by the RINA for exhaust monitoring as well as other applications where continuous monitoring of gasses is relevant. It can be installed as CEMS on SO2 scrubbers according to MEPC.340(77). The system can also be applied as a generic gas monitor for various gasses. For these applications the mentioned relevant sections of NTC 2008, appendix 3 have been demonstrated. The G7200 has been evaluated and found fit for purpose for dual fuel LNG engines in relation to the IGF code.
Acceptance Condition:
Correct configuration and set up for each delivery to be tested during commissioning after installation.
The G7200 Multi Gas Monitoring System shall be installed, maintained and operated in compliance with the User Manual.
Remarks:
This Certificate has replaced the previously issued Certificate No. MAC100123XP & MAC045724XP.